High quality 13 frequency DPgrams with settable normative range markers

DP growth I/O function analysis, dB/dB and scissor modes. Intelligent or manual test progress management for fast testing

Detailed numerical data available with each test result. Stimulus, response noise levels plus signal to noise ration SNR

DP zoom focuses on a specific range of frequencies in high resolution (up to 16 pts/oct) to explore details

Comprehensive TEOAE analysis. , high resolution spectrum

View TEOAE waveform

View TEOAE signal and noise histogram

View TEOAE numerical data

In-Ear Calibration ensuring accurate results

Multi Test macro (x4 protocals) function- build a protocol giving a sequence of DP and/or TE tests

Comprehensive automatic or manual test setup of parameter, protocol and individual clinical pass criteria

Configurable as a simple-to-use TEOAE automated screener for hearing screening applications
APPLICATION

Wireless Connection: Included


Docking Station: Included
PAEDIATRIC DISPLAY: Included


ILO v6 SOFTWARE: Included
OTOLINK SUITE: Included


OTOPORT DESIGN
|
|
EASY FIT PROBE DESIGN

Accessories Included

UGS TEOAE Probe

Probe pouch

Probe Clip

UGD DPOAE+TE probe
Specification Table
SPEC | DPOAE | TEOAE |
---|---|---|
TEST CONFIGURABILITY | 7 Protocols, 16 Parameters | 3 Protocols, 16 Parameters |
FREQ RANGE (kHz) | 1 to 8 kHz, 13 Frequencies | 1, 1.5, 3, 4, 5, 6 kHz |
FREQ. RESOLUTION | Up to 16 pts/octave | 80Hz or 1/2 Octave bandwidth |
STIMULUS LEVELS (dBSPL) | 40 to 75 (Δ5dB) | 60 to 90 (Δ1db) |
PASS CRITERIA | 9 Parameters | 7 Parameters |
PASS SNR | 1 to 14dB per band (Δ1dB) | 1 to 14dB (Δ1dB) |
Additional Options

ADD AABR

ADD NOAH