Easy-to-use one key screening or clinical setup

Link test macro function - build a protocol giving a sequence of DP and/or TE tests

In-ear calibration ensuring accurate results

High quality 12 frequency line DPgrams or higher resolution with normative range markers from 1-8kHz up to 4pts/octave and or higher resolution

High quality 12 frequency bar DPgrams or higher resolution with normative range markers from 1-8kHz up to 4pts/octave and or higher resolution

Numerical table of signal, noise and SNR

Any 2, 3, 4 or 5 bands configurable DP pass criteria

Live TEOAE waveform

CLINICAL AUDIOLOGY
PAEDIATRIC ASSESSMENT


NEWBORN SCREENING
PAEDIATRIC DISPLAY: Included


OTOLINK SUITE: Included
Manage your data easily and efficiently with the Otolink PC patient data management suite. Download test results wirelessly* or via USB from the Otoport or when docked in the docking station**. Test details can be quickly reviewed, printed and archived. Printing reports directly to PDF file is available. Data can be exported, for example, to Oz, Hi Track, GDT, Noah*** or as CSV file. Patient work lists can be uploaded to the Otoport, assisting with managing workloads in busy environments. Otolink also provides access to firmware and application updates from Otodynamics via the internet.
Compatible with Windows 10, 8 and 7.
LANGUAGE OPTION
Set the language you require through the Otoports configuration menu
|
|


OTOPORT DESIGN
|
|
EASY FIT PROBE DESIGN

Accessories Included

UGS TEOAE Probe

Probe pouch

Probe Clip

UGD DPOAE+TE probe
Specification Table
SPEC | DPOAE | TEOAE |
---|---|---|
TEST CONFIGURABILITY | 6 Protocols, 11 Parameters | 2 Protocols, 10 Parameters |
FREQ RANGE (kHz) | 1 to 8 kHz, 12 Frequencies | 1, 1.5, 2, 3, 4kHz |
FREQ. RESOLUTION | Up to 4 pts/octave | 1/2 Octave bandwidth |
STIMULUS LEVELS (dBSPL) | 60/50 to 70/70 (6 Presets) | 60 to 84 (Δ1db) |
PASS CRITERIA | 3 Parameters | 2, 3, or 4 1/2 octave, wide or narrow bands to pass |
PASS SNR | 1 to 10dB per band (Δ1dB) | 6dB (Fixed) |